3
Advanced Maintenance
Replacing Components of the Vacuum Interlock
232
ISQ 7000 Hardware Manual
Thermo Scientific
27. Reattach the two screws connecting the shaft of the vacuum interlock handle to the valve
body.
28. Slide the pillow block over and onto the end of the valve axle.
Note
You need to fully close the sealing ball with the valve axle before you put the handle
back on. The handle in the fully down position when you clamp it to the valve axle.
Figure 235.
Reattaching the Vacuum Interlock Handle to the Valve Body
Pillow
Block
Valve Axle
Screws
Vacuum
Interlock
Handle Shaft