Advanced Maintenance
Replacing the Covers of the ISQ 7000 Mass Spectrometer
280
ISQ 7000 Hardware Manual
Thermo Scientific
Top Manifold Cover
To replace the top manifold cover
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. Remove and replace the glass manifold cover on the top of the instrument.
4. Reattach the top cover panel and tighten the screw holding it in place.
5. Restart your system by following the instructions in
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
Figure 294.
Replacing the Top Manifold Cover
Top Manifold Cover