Advanced Maintenance
Replacing the Covers of the ISQ 7000 Mass Spectrometer
Thermo Scientific
ISQ 7000 Hardware Manual
281
Replacing the Chassis Feet
To replace the chassis feet on the bottom of the ISQ 7000 instrument
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the ISQ 7000 Instrument”
3. Follow the instructions in the section,
to remove the
components required to safely replace the chassis feet.
4. Position the instrument so that one foot hangs off the edge of the workbench. You may
have to move the instrument around on the workbench to replace all four feet.
5. Use a T20 Torxhead screwdriver to remove the four screws holding any damaged feet to
the bottom of the instrument.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering this
component.
Note
If you cannot hang the feet off the edge of the workbench, you can lay the
instrument on its left side and replace the feet. However, you will have to remove the top
cover panel and glass manifold cover first.
Figure 295.
Removing the Feet on the Bottom of the ISQ 7000 Instrument
Feet