Performing Routine Maintenance
Cleaning the Ion Source Cartridge
70
ISQ 7000 Hardware Manual
Thermo Scientific
5. With the longer teeth of lens 1 facing down toward lens 2, align the larger metal section
of lens 1 with the lens spring contact slot and let it fall into place.
6. Insert the ion volume with the alignment wings fitting into the notches of the sleeve.
Make sure the alignment tooth ion volume is firmly seated into the gap on lens 1.
Figure 62.
Inserting Lens 1 into the Ion Cartridge Sleeve
Figure 63.
Inserting the Ion Volume into the Ion Cartridge Sleeve
Longer Teeth
of Lens 1
Larger Metal
Section
Transfer Line and Lens
Spring Contact Slot
Alignment
Wings
Lens 1 Gap
Notches
Alignment Tooth