1
Performing Routine Maintenance
Maintaining the Foreline Pump
Thermo Scientific
ISQ 7000 Hardware Manual
45
.
27. Open the front door of the instrument.
28. Look behind the right side of the vacuum interlock shield and twist the vent valve knob
one and a half times in a counter-clockwise direction to open the vent.
29. Wait five minutes for venting to complete.
Figure 30.
Powering Off the ISQ 7000 Instrument
Power Switch
Vacuum Interlock Shield
Vent Valve
Knob
Hydrogen Safety Screws
Hydrogen Safety Screw
(On Front Door Panel)
CAUTION - INSTRUMENT DAMAGE
Do not proceed until the instrument is vented, or
pieces of the column or ferrule might blow into the instrument. To ensure that the
instrument is vented, check how much the glass cover compresses the top cover o-ring in
the manifold. Once the o-ring surface touching the glass is about 1 mm, it is safe to open
the instrument and remove the column.
ATTENTION DOMMAGES À L’INSTRUMENT
Ne pas poursuivre tant que l’instrument n’a
pas été ventilé, ou des morceaux de la colonne ou de la virole pourraient éclater dans
l’instrument. Pour vous assurer que l’instrument est ventilé, vérifiez dans quelle mesure le
couvercle en verre comprime le joint torique du couvercle supérieur dans le collecteur.
Lorsque la surface du joint torique en contact avec le verre équivaut à environ 1 mm, vous
pouvez sans danger ouvrir l’instrument et enlever la colonne.
Содержание ISQ 7000
Страница 1: ...ISQ 7000 Mass Spectrometers Hardware Manual 1R120617 0002 Revision C December 2018 ...
Страница 8: ......
Страница 174: ...2 Troubleshooting Contacting Technical Support 146 ISQ 7000 Hardware Manual Thermo Scientific ...
Страница 400: ...3 Advanced Maintenance Replacing Fans and Filters 372 ISQ 7000 Hardware Manual Thermo Scientific ...