3
Advanced Maintenance
Replacing Components of the Analyzer
Thermo Scientific
ISQ 7000 Hardware Manual
155
Figure 121.
Identifying the Set Screw on the Source Heather Block
14. Remove this screw with a T10 screwdriver. It is important to remove this screw
completely so that it cannot damage the transfer line during disassembly/reassembly.
Place this screw in a clean place until it is needed for reassembly. See
Figure 122.
Removing the Set Screw
Set Screw
Source Interface
Board
Set
Screw
Calibration and
Reagent Gas
Tubes
Right Screw