CI Reagent Gas Flow Module
The CI reagent gas flow module controls the flow of CI reagent gas into the ion source and is
located at the back left of the ISQ 7000 mass spectrometer. The module uses electronic
pressure control to precisely set the flow of reagent gas. The flow is programmed using
Xcalibur
, which allows you to accurately reproduce flows. To install a CI reagent gas flow
module:
1. Shut down your system by following the steps in
“Shutting Down the ISQ System”
2. Use a T20 Torxhead screwdriver to loosen the top cover panel and left side panel screws
from the interior front panel. Then slide the left and top cover panels toward the back of
the instrument and lift them off.
3. If the ISQ 7000 mass spectrometer is too close to the GC, move them apart so you can
access the right side panel.
4. Use a T20 Torxhead screwdriver to remove the screw at each corner of the right side
panel.
5. Slide the right side panel toward the back of the instrument and lift it off.
Figure 435.
Removing the Right Side Panel Screws
Right Side Panel Screws