3
Advanced Maintenance
Replacing Components of the Analyzer
172
ISQ 7000 Hardware Manual
Thermo Scientific
Ion Guide Wire Set
To replace the ion guide wires:
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Remove the analyzer tray from the instrument by following the instructions in
“Removing the Analyzer Tray from the ISQ 7000 Instrument”
3. Use a T10 Torxhead screwdriver to remove the screw connecting the ion guide wire to the
top of the ion guide.
4. Use a wrench to remove the nut connecting the other ion guide wire to the side of the ion
guide.
5. Attach new ion guide wires and reattach the nut and screw.
6. Replace the analyzer tray by following the instructions in
.
7. Restart your system by following the instructions in
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components.
Figure 144.
Replacing the Ion Guide Wires
Screw
Nut
Ion Guide Wires