3
Advanced Maintenance
Replacing Components of the Analyzer
158
ISQ 7000 Hardware Manual
Thermo Scientific
23. Move the large tab on the analyzer tray to the left, moving the ion source away from the
transfer line, and then carefully pull the tray toward you through the manifold door.
24. Set the tray on a clean work surface.
CAUTION - INSTRUMENT DAMAGE
Do not scratch any of the sealing surfaces. Be
especially careful when the source interface board reaches the o-ring surface.
ATTENTION DOMMAGES À L’INSTRUMENT
Ne rayez pas les surfaces d’étanchéité. Soyez
particulièrement prudent lorsque la carte d’interface de la source entre en contact avec la
surface du joint torique.
Figure 126.
Sliding the Analyzer Tray Out of the Instrument
Analyzer Tray
Tab