Advanced Maintenance
Replacing Components of the Vacuum Interlock
Thermo Scientific
ISQ 7000 Hardware Manual
223
To replace the components of the vacuum interlock
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Remove the vacuum interlock knob and use a T10 Torxhead screwdriver to remove the
five screws around the knob. Replace the knob and screws, if necessary.
Note
You do not have to remove the ion source cartridge or the column. However, you
must vent the instrument.
Note
As you remove the screws around the vacuum interlock knob, watch for a small
spring and safety interlock (location shown below) that will fall out as your remove the
shield. Be sure to catch them and set them aside. Replace the spring and safety interlock, if
necessary.
Figure 219.
Replacing the Vacuum Interlock Knob
Screws
Vacuum
Interlock Shield
Spring and safety interlock
are behind this screw