Electrometer Board and Shield
To replace the electrometer board and shield
1. Shut down your system by following the steps in
“Shutting Down the ISQ 7000 System”
.
2. Use a T20 Torxhead screwdriver to loosen the top cover panel screw from the interior
front panel. Then slide the top cover panel toward the back of the instrument and lift it
off.
3. Disconnect the electrometer board cable.
Note
See the
ISQ and TSQ GC-MS Spare Parts Guide
for information about ordering
these components.
Note
You do not have to remove the ion source cartridge or the column or vent the
instrument. If you are using hydrogen as a carrier gas, power off the GC.
Figure 383.
Disconnecting the Electrometer Board Cable from the Electron Multiplier
Electrometer
Board Cable
Electron
Multiplier Plate
Electrometer
Shield