3
Advanced Maintenance
Replacing Components of the Manifold
Thermo Scientific
ISQ 7000 Hardware Manual
253
3. Open the manifold door and replace the front door alignment pins, magnet yoke, source
insertion guide, source magnets, or front manifold door o-ring as needed on the back of
the door,
4. Close the manifold door and reattach the four screws.
5. Restart your system by following the instructions in
Figure 264.
Replacing the Source Retainer Magnets, Bracket, and Front Manifold O-ring
Figure 265.
Closing the Manifold Door
Source Retainer
Magnets
Magnet
Yoke
Front Manifold
O-ring
Source
Magnets
Source
Insertion
Guide
Screws