Troubleshooting
Investigating Tuning Issues
144
ISQ 7000 Hardware Manual
Thermo Scientific
Cannot perform mass calibration.
An air leak has been detected.
Possible Remedies
Allow contaminants to pump away so they do not interfere with calibration gas ions.
Create a starting tune file using manual tune. Select Air & Water /Tune from the ISQ 7000 Dashboard to launch ISQ Manual Tune
Possible Remedies
Check the system for air leaks and address them.
Use a spring-loaded transfer line nut (PN 1R120434-0010) to prevent leaks with graphite-Vespel ferrules.