3
Advanced Maintenance
Replacing Components of the Electron Multiplier
Thermo Scientific
ISQ 7000 Hardware Manual
261
9. Flip the electron multiplier and plate over so that the plate is resting on the worktable.
10. Remove the two screws connecting the electron multiplier to the plate and attach the
electron multiplier to a new plate.
Figure 275.
Resting the Electron Multiplier Plate on the Worktable
Figure 276.
Replacing the Electron Multiplier Plate
Electron
Multiplier
Plate
Screws