Performing Routine Maintenance
Cleaning the Ion Source Cartridge
82
ISQ 7000 Hardware Manual
Thermo Scientific
Reassembling the Advanced EI Ion Source Cartridge
The ion source parts are assembled the same way as in
up until Lens 2, although you may use gloved hands instead of the
source holder.
1. Slide the AEI ion volume insulator over the AEI ion volume. Be sure to align the notch in
the insulator with the step on the ion volume. See
Figure 78.
Inserting the AEI Ion Volume Insulator Over the AEI Ion Volume
2. Insert the AEI ion volume with insulator into the Advanced EI ion source cartridge. Be
sure that the opening for the column is exposed in the slot of the Advanced EI ion source
cartridge. See