XK3118K9 Weighing Indicator Calibration Manual 

I. Main Technical Parameters: 

Load cell excitation:

 DC +5V, can connect 1

12 pieces of 350 ohm strain gauge type load cells 

Minimum input voltage of each calibration division value: ≥1.5uV/e         

Error Distribution coefficient: p

=0.5 

Load cell connection mode:

 6 wires (long wire auto compensation), compensation distance ≤50 meter 

Power supply:

 AC 110-220V, 50-60Hz,   

DC 6V/10Ah free maintenance lead-acid battery, can be charged by the indicator 

 
II. Analog Load Cell Connection: 

 

Specification 

Pins 

Name   

Code 

Excitation - 

E- 

Feedback - 

F- 

Exci 

E+ 

Fe 

F+ 

8

 

Signal - 

S- 

 

S+ 

Shield 

SHIELD 

E+ with F+, E- with F- must be short connected if you use 4 wire shield cable. 

 

III. Calibration 

Indicator enters weighing status after self-check. When calibration is needed, open the cover of calibration case on the 

back side of indicator, press the calibration touch switch. Then the beep sounds and calibration lamp lights up, which 

indicates  that  the  calibration  mode  can  be  entered.  Then  press  the  calibration  button  on  the  front  panel.  Then  the 

calibration lamp goes out and enters calibration mode. Detailed calibration steps are as below: 

Step 

Operation 

Display 

Remarks 

Press

Calibration

 

d      *

 

After  calibration  is  finished  or  interrupted,  if  need  to  calibrate  again, 
the calibration switch should be pressed one more time. 

Press

1

0

          Press

Input

 

d      10

 

Input  division,  1/2/5/10/20/50/100  optional,

  (factory  default  setting: 

10) 
If need compensation calibration, press 

0

press

Input

enters 

to  step  7,  indicator  displays 

AdloAd1

,  input  weight  value  of 

weights loaded, press   

Input

to confirm 

 

Press

  0 

            Press

Input

 

dC    0 

 

Input decimal bits ( 0

4 ), default setting: no decimal 0 

Press

1

2

2

4

 

Press

Input

 

Pn 01224

 

0 1 2 2 4 

Corresponding 

VWXYZ 

 

Input system parameter (details reference to  the table below)

  (factory 

default setting: 01224)

 

V: application (0: non-business scale 1: business scale) 
W: stability range 
X: zero tracing range 
Y

manual zero-setting range 

Z

power on zero-setting range 

Press

3

0

0

0

0

 

Press

Input

 

F 30000

 

Input full capacity value 

(factory default setting:3000)

 

If need calibration, input F value

then press 

Input

;if press

input

directly  then  it  enters  to  step  8

if  press

weight

,return  back  to 

weighing status. 

note

full capacity value add 9 division value equals to system preset 

overload alarming value.

 

Press

Input

 

  noLoAd

 

Zero position confirmation

no object on the scale, press 

Input

after 

stability symbol light becomes bright 

 

★★★

!!!

    When indicator displays 

noLoAd

press 

Weigh

button, skip zero calibration, enters to 

compensation calibration directly 

Press

2

0

0

0

0

 

Press

Input

 

AdloAd1

 

    20000

 

Add weights, not less than 50% of full capacity, more accurate when it 
is near full capacity. 
e.g.

20000 