Test Results
Power Blazer
579
Summary (S-OAM and MPLS-TP OAM)
Continuity Check (Peer MEP)
Status
displays the status of the continuity check with the peer MEP.
TX CCM
indicates the count of transmitted CCM frames.
RX CCM
indicates the count of valid CCM frames received. A valid
frame for S-OAM has its source MAC address matching the Peer MEP
MAC address; destination MAC address matching either the unit port
Unicast MAC address or a Multicast class 1 address (refer to 841); and
VLANs matching the unit port VLANs. A valid frame for MPLS-TP OAM
has its destination MAC address matching either the unit MAC address,
FF:FF:FF:FF:FF:FF, or 01:00:5E:90:00:00; VLANs matching the unit port
VLANs; and MPL Labels matching the local MPLS Label Stack
configuration.
Status
Description
Loss Continuity
Loss of Continuity alarm is active.
Mismerge
Mismerge alarm is active.
Unexpected MEG Level Unexpected MEG Level alarm is active.
Unexpected MEP
Unexpected MEP alarm is active.
Unexpected Period
Unexpected Period alarm is active.
Unexpected MD Level
Unexpected MD Level alarm is active.
Receiving CCMs
CCM frames from the peer MEP are received without alarms.
Summary of Contents for Power Blazer 88200NGE
Page 1: ...www EXFO com User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE...
Page 14: ......
Page 20: ......
Page 64: ...Test Setup Test Applications 54 88000 Series OTN BERT...
Page 71: ...Test Setup Test Applications Power Blazer 61 SONET SDH BERT For SDH BERT...
Page 76: ...Test Setup Test Applications 66 88000 Series SONET SDH DSn PDH BERT For SDH...
Page 620: ......
Page 726: ......
Page 752: ......
Page 756: ......
Page 768: ......
Page 886: ...0 5 1 5 48 5 0 176 W d y K y K D...