Test Setup - Test Configurator, Timer, and System
218
88000 Series
FTFL/PT and PT
FTFL
Allows the configuration of the Forward and Backward ODU Fault Type
Fault Location (FTFL) to be generated. Not available with Multi-Channel
OTN.
Overwrite
check box when selected generates the defined FTFL. Only
available with Through Intrusive topology and applies only to the ODU
top layer.
Fault Indication
and
Fault Indication Code
allow the selection of the
FTFL fault indicator message/code (byte 0 for forward, byte 128 for
backward) to be generated.
Note:
The
Fault Indication Code
field is automatically updated when the
Fault
Indication
is changed and vice versa.
Operator Identifier
allows editing the Operator Identifier to be
generated (bytes 1 to 9 for forward, bytes 129 to 137 for backward;
9 characters allowed). By default no Operator Identifier is defined.
Operator Specific
allows editing the Operator Specific to be
generated (bytes 10 to 127 for forward, bytes 138 to 255 for backward;
118 characters allowed). By default no Operator Specific is defined.
Fault Indication
Fault Indication Code (Hex)
No fault
00
(default)
Signal fail
01
Signal Degrade
02
Reserved
03
a
a.
Selecting
Reserved
will use the hexadecimal code
03
but, all codes from
03
to
FF
are reserved for future international standardization.
Summary of Contents for Power Blazer 88200NGE
Page 1: ...www EXFO com User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE...
Page 14: ......
Page 20: ......
Page 64: ...Test Setup Test Applications 54 88000 Series OTN BERT...
Page 71: ...Test Setup Test Applications Power Blazer 61 SONET SDH BERT For SDH BERT...
Page 76: ...Test Setup Test Applications 66 88000 Series SONET SDH DSn PDH BERT For SDH...
Page 620: ......
Page 726: ......
Page 752: ......
Page 756: ......
Page 768: ......
Page 886: ...0 5 1 5 48 5 0 176 W d y K y K D...