Test Setup - Test Configurator, Timer, and System
Power Blazer
189
EtherBERT, FC BERT, BERT (eCPRI, CPRI/OBSAI, and FlexE), and Unframed BERT
Ethernet Frame
Note:
For eCPRI see eCPRI Flow - Profile
.
Frame Size (Bytes)
allows entering the frame size for Ethernet test
applications. The range is from
64
1
to
16000
2
. EMIX is only available for
rates up to 100GE.
Fixed
(default)
The following table lists each component that may affect the
minimum frame size values.
Note:
Sending traffic with frame size >1518 in switched network may result in
losing all frames.
EMIX
button is available when EMIX frame size is selected. The
EMIX frame sequence is repeated continuously.
1. The minimum frame size will be adjusted according to the frame structure and components selected.
2. The maximum frame size is limited to 10000 for 10/100/1000Mbps electrical interface.
Component
Description
VLAN
4 bytes per VLAN (up to 3 VLAN for EtherBERT test
application and only 1 VLAN for EoOTN client)
EoE Header
16 bytes
EoE VLAN
4 bytes
PBB-TE Header
18 bytes
B-VLAN 4
bytes
UDP
8 bytes
TCP 20
bytes
Ethernet Header
14 bytes
LLC and SNAP Headers 8 bytes
IPv4
20 bytes
IPv6 40
bytes
Summary of Contents for Power Blazer 88200NGE
Page 1: ...www EXFO com User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE...
Page 14: ......
Page 20: ......
Page 64: ...Test Setup Test Applications 54 88000 Series OTN BERT...
Page 71: ...Test Setup Test Applications Power Blazer 61 SONET SDH BERT For SDH BERT...
Page 76: ...Test Setup Test Applications 66 88000 Series SONET SDH DSn PDH BERT For SDH...
Page 620: ......
Page 726: ......
Page 752: ......
Page 756: ......
Page 768: ......
Page 886: ...0 5 1 5 48 5 0 176 W d y K y K D...