Test Setup - Test Configurator, Timer, and System
Power Blazer
281
ODU Channels - Global
Defect
allows choosing on which layer and defect the service
disruption time test will be performed. Choices depend on the
selected test path.
Note:
The Service Disruption Time measurement supports a parent defect
approach where the SDT measurement is triggered when the selected
defect or a higher defect in the signal structure hierarchy is detected.
No Defect Time (ms)
represents the period without any defects before
stopping SDT measurement:
0.005 ms
to
2000 ms
(default is
300 ms
).
Pass/Fail Verdict
check box when selected (cleared by default)
enables service disruption pass/fail verdict and allows setting the
threshold value.
SDT Threshold (ms)
allows entering the SDT threshold value that will
be used to declare the pass/fail verdict:
0.001
to
299999.999 ms
(default is
50
ms).
Layer
Defect
ODUk
LOFLOM, AIS
,
OCI
,
LCK
,
BDI
,
BIP-8
,
BEI
OPUk
AIS
,
CSF
Summary of Contents for Power Blazer 88200NGE
Page 1: ...www EXFO com User Guide 88000 Series Power Blazer HIGH SPEED MULTISERVICE TEST MODULE...
Page 14: ......
Page 20: ......
Page 64: ...Test Setup Test Applications 54 88000 Series OTN BERT...
Page 71: ...Test Setup Test Applications Power Blazer 61 SONET SDH BERT For SDH BERT...
Page 76: ...Test Setup Test Applications 66 88000 Series SONET SDH DSn PDH BERT For SDH...
Page 620: ......
Page 726: ......
Page 752: ......
Page 756: ......
Page 768: ......
Page 886: ...0 5 1 5 48 5 0 176 W d y K y K D...