Powering On the System
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
3–23
Circuit breaker
The Circuit Breaker is located on the back of the console, at the
bottom of the system. On supplies main power to all internal
systems. Off removes main power from all internal systems.
The circuit breaker automatically shuts off power to the system
in case of a power overload.
If a power overload occurs:
1.
Turn off all peripheral devices.
2.
Turn off the Main Power Switch to the console.
3.
Reactivate the Circuit Breaker switch.
The Circuit Breaker switch should stay in the On position; DO
NOT hold the switch in the On position. If the Circuit Breaker
switch remains On, follow the
Power On
procedure previously
described.
NOTE: If the Circuit Breaker switch does not remain in the On
position or trips again:
1.
Disconnect the Power Cable.
2.
Call Service immediately.
DO NOT attempt to use the system.