Recording Images
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
13–22
Laser Camera
The LOGIQ
500 can print images to a Laser Camera for
archival.
Contact the local GE Service Representative for details about
hook-up and operation.
Video Signal Specifications
Table 13–2 gives the video specifications of the Ultrasound
system. This data may be needed when interfacing the system
with OEM cameras or recording devices.
NTSC
PAL
Total Lines per Frame [line]
525
625
Vertical Field Frequency [Hz]
60
50
Horizontal Scanning Frequency [Hz]
15.733
15.625
Displayed Image Pixels [mm]
207.0 by 157.3
207.0 by 157.3
Total Horizontal Line Time [ms]
63.56
64.00
Horizontal Display [ms]
49.54
48.81
Front Porch Width [ms]
2.76
3.09
Sync Pulse Width [ms]
4.73
4.68
Back Porch Width [ms]
6.53
7.42
Total Horizontal Blanking [ms]
14.02
15.19
Vertical Blanking Interval [H]
31.50
38.50
Vertical Front Porch Width [H]
6.5
9.0
Vertical Sync Width [H]
3
2.5
Vertical Back Porch Width [H]
22
27
Table 13–2. Video Signal Specifications
Maintenance
Consult the peripheral device operator manual for necessary
routine maintenance.