LV Measurement Methods
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
10–16
Single Plane Ellipsoid Method (cont’d)
Calc Mnemonic
Calc Name
Input Measurements
Formula
LVLd
Left Ventricular Length,
Diastole
one distance
LVLd=d1[cm or mm]
LVLs
Left Ventricular Length,
Systole
one distance
LVLs=d1[cm or mm]
LVAd
Left Ventricular Area,
Diastole
one area (by ellipse,
trace or circle)
LVAd=a1[cm ^2]
LVAs
Left Ventricular Area,
Systole
one area (by ellipse,
trace or circle)
LVAs=a1[cm ^2]
EdV
End Diastole Volume
one distance and two
areas (by ellipse, trace
or circle)
EdV[ml]=8/(3
p
)
x(LVAd)^2/LVLd
EsV
End Systole Volume
one distance and two
areas (by ellipse, trace
or circle)
EsV[ml]=8/(3
p
)
x(LVAs)^2/LVLs
SV
Stroke Volume
two distances and four
areas (by ellipse, trace
or circle)
SV[ml]=EdV–EsV
CO
Cardiac Output
two distances and four
areas (by ellipse, trace
or circle) and one 2 beat
time interval
CO[1/min]=SVxHR/1000
EF
Ejection Fraction
two distances and four
areas (by ellipse, trace
or circle)
EF=SV/EdVx100
HR
Heart Rate
(beats/minute)
one 2 beat time interval
HR[BPM]=120 [sec]/ 2
beat time [sec]
PHT
Pressure Half Time
one time interval
PHT=t1[ms orsec]
MVA
Mitral Valve Area
one pressure half time
interval
MVA[cm^2]= 220/PHT
ET
Ejection Time
one time interval
ET=t1[ms or sec]
Table 10–5. Single Plane Ellipsoid Method Formulas
Summary of Contents for 500
Page 1: ... ...