Custom Display
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
14–30
Page 9 of 20 (Imaging Parameter 9 – Probe Dependent 9) (cont’d)
PFD Packet Size EC, UP, P
Choose the default packet size for each PFD diagnostic mode display.
Small
Small packet size.
Mid
Medium packet size.
Large
Large packet size.
NOTE: The larger the packet size the better the color presentation but with a slower
frame rate.
PFD Scale EC, UP, P, R
Choose the default PFD velocity scale in centimeters per second.
Value from 1 to 27 in 1 digit increments.
PFD B Gain EC, UP, P, R
Choose the default B-Mode gain setting while in PFD Mode.
Value from 0 to 31 in 1 digit increments.
PFD B High Resolution EC, UP, P, R
Choose the initial default setting of the high resolution function in PFD Mode.
On or Off.
PFD M Gain (Delta from B) EC, UP, P, R
Choose the default level for M-Mode gain while in PFD Mode.
Level from –31 to 31 in 1 digit increments.
PFD M Wall Filter EC, UP, P, R
Choose the default level for the M-Mode Wall Filter while in PFD Mode.
Low, M1, M2, M3, M4, or High.
PFD ACE EC, UP, P, R
Choose the default setting for the PFD Adaptive Color Enhancement, which reduces clutter
noise or color flash.
ON or OFF.
PFD Resolution EC, UP, P, R
Choose the default setting for the PFD resolution level.
1A, 2A, 3A, 1B, 2B, or 3B
PFD Persistence EC, UP, P, R
Choose the default setting for the PFD Persistence, which retains the highest color pixel
value for a specified period of time or until a higher value is detected.
Off, 1, 2, 3, 4, 5 or 6