Custom Display
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
14–26
Page 7 of 20 (Imaging Parameter 7 – Probe Dependent 7)
CFM M Gain (Delta from B) EC, UP, P, R
Choose the default level for M-Mode gain while in CFM.
Level from –31 to 31 in 1 digit increments.
CFM M Wall Filter EC, UP, P, R
Choose the default level for the M-Mode Wall Filter while in CFM Mode.
Low, M1, M2, M3, M4, or High.
CFM ACE EC, UP, P, R
Choose the default setting for Adaptive Color Enhancement, which reduces clutter noise
or color flash.
On or Off.
CFM Noise Blanker EC, UP, P, R
Choose the default setting for Noise Blanker, which reduces random color noise.
OFF, LOW, MID or HIGH.
CFM Persistence EC, UP, P, R
Choose the default setting for Persistence, which retains highest color pixel value for a
specified period of time or until a higher value is detected.
OFF, SHORT, MID or HIGH.
CFM MR-Flow EC, UP, P, R
Choose the default setting of the MR-Flow option (combination focus in CFM area).
Limited to LA39, C551 (CAE), C358, E721, 739L and 546L probes only. Must have
ACE-2 option installed.
On or Off.
CFM MR-Flow Combi Number EC, UP, P
Set the number of focal zones to be used for the combination focus for MR-Flow option.
2 or 3.
CFM Map Transparency EC, UP, P
Choose the default CFM Transparency map. The higher the number, the more
transparent the map.
Off, 1, 2, 3, or 4.
PDI Map Transparency EC, UP, P
Choose the default PDI Transparency map. The higher the number, the more
transparent the map.
Off, 1, 2, 3, or 4.