Preparing for a Biopsy
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
15–45
Multi-Angle Biopsy Guide Assembly
DO NOT attempt to use the multi-angle biopsy bracket and
needle guide until the manufacturer’s instructions, provided with
the biopsy bracket and needle guide in the kit, have been read
and thoroughly understood.
Scan the patient and identify the target for biopsy. Move the
probe to locate the target to the center of the image. Enable
the system biopsy guidezone and try guidezone angles MBX1
to MBX3 to decide the best angle setting for needle path.
NOTE: The biospy guidezone type can be selected in the Set
Up/Custom Display menu page 1.
Identify the appropriate biopsy guide bracket by matching the
label on the bracket with the probe to be used.
Î
Figure 15–22. S317 Multi-Angle Biopsy Guide Bracket
Orient the bracket so that the needle clip attachment will be on
the same side as the probe orientation mark (ridge).
1
2
Î
1
Probe Orientation Mark
2
Bracket
Figure 15–23. Probe/Bracket Alignment
Attach the biopsy bracket to the probe by sliding the bracket
over the end of the probe until it clicks or locks in place.
WARNING