Troubleshooting
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
16–29
Operation Error Message Description (cont’d)
Category
Problem/
Message
Possible
Cause
Possible
Corrective Action
Recording
Devices
M l i I
Check M.I.C., No
Exposures Left.
The maximum number of
prints have been made.
Reverse cassette or
install a new cassette.
Multi Image
Camera
Check M.I.C., No Video
Input.
There is no video signal
being input to the M.I.C.
Check all video
connections.
Check M.I.C.
There is no response
from the M.I.C.
Turn power off, turn
power on.
Recording
Devices
VCR
Check VCR.
There is no response
from the VCR.
Check all VCR controls.
VCR
Check VCR. No
Cassette.
The VCR has no tape.
Insert tape.
This function is not
available.
The function selected is
not available in some
situations.
Perform from VCR front
panel.
Protected. Check Tape.
This tape has been write
protected.
Use non protected tape.
Line Printer
Check Line Printer. The
port isn’t set.
The line printer has not
been assigned to Port
MIC, A or B.
Assign the line printer to
Port MIC, A or B.
Check Line Printer. No
Power.
The line printer is
powered off or the serial
cable is not connected to
port MIC, A or B.
Power on the line printer
or connect the serial
cable to port MIC, A
or B.
Check Line Printer.
Printing Error.
An error has occurred in
sending data to the line
printer or the printer is
out of paper.
Resend the report data
or add paper to the
printer.
Check Line Printer.
Printing isn’t finished.
Report Print Buffer is full.
Ctrl + R cannot be used
until space is available.
The system can store 10
reports for background
printing.
Table 16–4. Operation Error Message Description (cont’d)
Summary of Contents for 500