Preparing for a Biopsy
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
15–37
Preparing for a Biopsy
Displaying the Guidezone
Activate the Biopsy Zone by selecting it from the B-Mode
Sub-Menu.
SGL
Preset
B
Edge
Enhance
Rejectn
Image
Rotatn
Biopsy
Zone
Set Up
ECG
0DEG
40
MID
Figure 15–10. B-Mode Sub-Menu (Biopsy Zone)
If the single fixed angle (SGL) was selected as the needle guide
type in Set Up/Custom Display page 1, the fixed zone angle is
displayed.
If the needle guide type selected is multiple angle (MULTI), the
Biopsy Zone rocker switch can be used to cycle through the
angle selections MBX1, MBX2, MBX3 and OFF.
The Biopsy Guidezone adjusts along with image adjustments,
such as image inversion/rotations, zoom and depth changes.
MBX3
Figure 15–11. Biopsy Guidezone Example