Operator Diagnostics
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
16–36
System Test 1 (reduced)
Select 01 from the diagnostic test menu.
System Test 1 is a collection of 24 individual tests that are run
in sequence. These are the full or reduced versions of the
hardware tests.
Three possible messages will be displayed to the right of the
test name.
RUNNING—The test is in progress.
FAILED—The test was not completed successfully for some
reason.
PASSED—The test was run and completed successfully.
Figure 16–11 shows the screen display during system Test 1
operation.
GE MEDICAL SYSTEMS
03 Beam Former Test 1
04 Beam Former Test 2
05 Log Test
06 Frequency Characteristic Test
08 Doppler Test 1
09 Doppler Test 2
11 Color Flow Test 1
12 Color Flow Test 2
13 TLMP Test (Reduced)
14 DDSC Test (Reduced)
15 CINE Test (Reduced)
16 VIDO Test (Reduced)
17 VIDO & VPBM Test (Reduced)
18 EXBO Test (Reduced)
19 RCTL Test (Reduced)
20 PIOP Test (Reduced)
21 Keyboard RAM Test (Reduced)
27 Power Supply & Thermal Test
29 B Mode Noise Floor Test
30 D Mode Noise Floor Test
31 CFM Mode Noise Floor Test
03/21/94 01: 31: 50
System Test
Running
Figure 16–11. System Test 1 Screen Display
Should any test fail, consider that the system may be operating
at reduced capabilities. Record the System Test 1 results and
contact a local service representative.
Total run time for System Test 1 is approximately 15 minutes.