Preset Program
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
14–74
Page 3 of 10 (Application – Measurement) (cont’d)
Diastole/Systole Determination EC
Used with Advanced Cardiac Calculation option. Determine if measurement is systole
or diastole automatically from the measurements relationship to the ECG wave.
Manual
Automatic determination disabled.
Auto
Automatic determination enabled.
R Delay Time of End Systole Prior Edge EC, UP
Used with automatic determination of systole/diastole in the Advanced Cardiac
Calculation option. Sets the delay time on the ECG wave to ensure systole will be
properly determined.
Use the Ellipse rocker switch to enter a number between 0 and 2000 msecs.
End Systole Period from Prior Edge EC, UP
Used with Automatic determination of Systole/Diastole in the Advanced Cardiac
Calculation option. Sets the delay time on the ECG wave to ensure diastole will be
properly determined.
Use the Ellipse rocker switch to enter a number between 0 and 2000 msecs.
Media Selection for Fetal Trend EC
Used with the Fetal Growth Trend option. Sets the media to be used to store patient
data. The hard drive has a much limited capacity.
HD
System Hard Drive.
MO
Magnetic Optical Disk (removeable).
Measurement Clear Operation EC
When Clear is pressed, decide if it will erase measurements only or measurements and
comments.
Meas.–only
Erase only measurements not comments.
with–Comment Erase measurements and comments simultaneously.