Powering On the System
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
3–22
Power Off
When switching off the system:
Move the ON, OFF/STAND-BY switch to the OFF position.
The message “WARNING: NOW STARTING THE POWER
OFF PROCESS” appears at the bottom of the display.
The LOGIQ
500 takes a few seconds to save current
scan parameter data in the temporary files to the hard drive
before turning the power off.
During this time a message flashes on the screen:
“Do not pull Power Cable.
Do not turn off Breaker.”
Pulling the power cable or turning off the circuit breaker,
while the hard drive is working, may corrupt the system
operating software on the hard drive.
If the system has not turned off five minutes after pressing
the power switch off, listen for hard drive activity. If there is
no hard disk drive activity, the circuit breaker on the bottom
of the power supply can be used to turn off the system. Do
NOT turn off the circuit breaker while the hard disk is
working.
Disconnect the probes.
Clean or sanitize all probes as necessary. Store them in
their shipping cases to avoid damage.
If daily maintenance is to be performed, turn off the circuit
breaker in the back of the system.