GYN Calculation Formulas
Calc
Mnemonic
Calc Name
Input
Measurements
Formula
UT-L
Uterine Length
one distance
Ut-L[cm or mm]=d1
UT-H
Uterine Height
one distance
Ut-H[cm or mm]=d1
UT-W
Uterine Width
one distance
Ut-W[cm or mm]=d1
Endo
Endometrium Thick-
ness
one distance
Endo[cm or mm]=d1
Lt. Ov-L
Left Ovarian Length
one distance
Lt. Ov-L[cm or mm]=d1
Lt. Ov-H
Left Ovarian Height
one distance
Lt. Ov-H[cm or mm]=d1
Lt. Ov-W
Left Ovarian Width
one distance
Lt. Ov-W[cm or mm]=d1
Rt. Ov-L
Right Ovarian Length
one distance
Rt. Ov-L[cm or mm]=d1
Rt. Ov-H
Right Ovarian Height
one distance
Rt. Ov-H[cm or mm]=d1
Rt. Ov-W
Right Ovarian Width
one distance
Rt. Ov-W[cm or mm]=d1
Lt. Ov-RI
Left Ovarian Vessel
Resistive Index
two Doppler blood flow
peak velocities
Lt. Ov-RI= (V
max
–V
diastole
)/V
max
*NOTE*
Ut-RI
Uterine Vessel Resis-
tive Index
two Doppler blood flow
peak velocities
Ut-RI= (V
max
–V
diastole
)/V
max
*NOTE*
Rt. Ov-RI
Right Ovarian Vessel
Resistive Index
two Doppler blood flow
peak velocities
Rt. Ov-RI= (V
max
–V
diastole
)/V
max
*NOTE*
LtOvFo[ml]
Left Ovary Follicles
One distance*
D1[cm]
3
x
p
/6
Two distances*
D1[cm]
2
x D2x[cm]
p
/6
: (D1 <D2)
D1[cm] x D2[cm]
2
x
p
/6
: (D2 <D1)
Three distances
D1[cm] x D2[cm]x D3[cm] x
p
/6
RtOvFo[ml]
Rith Ovary Follicles
One distance*
D1[cm]
3
x
p
/6
Two distances*
D1[cm]
2
x D2[cm] x
p
/6
: (D1 <D2)
D1[cm] x D2[cm]
2
x
p
/6
: (D2 <D1)
Three distances
D1[cm] x D2[cm] x D3[cm] x
p
/6
*NOTE*:
V
diastole
= V
min
or V
end–diastole
(depends on preset selection)
* To calculate LtOvFo or RtOvFo with one (or two distances), press the Clear key
after the first (or second distance) measurement(s).
Table 9–8. GYN Calculation Formulas