Adding Color
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
5–47
PFD (Pulsatile Flow Detection) Mode (option)
Pulsatile Flow Detection Mode (option)
Description
PFD Mode shows pulsation of a flow superimposed on a PDI or Directional PDI image.
Flow existence, flow direction and vasculature information will be available.
NOTE: 3D images cannot be rendered in PFD Mode.
Accessing/
Changing
To activate PFD Mode:
While scanning in CFM or PDI mode, enable the Blue Shift key (key will be lit) and press P.
NOTE: PFD Mode is only operational with Power Doppler installed.
PFD Mode can also be activated by pressing the CFM key, if the CFM/PDI/PFD Initial
Mode preset, found on Custom Display page 18, is set to PFD. The CFM Mode
Sub-Menu pages change when PFD Mode is active.
The Slant Scan, Packet Size, Penetration and High Resolution parameters remain the
same. PFD Map Resolution, Persistence and PFD type reflect PFD Mode.
Values
“PFD” is displayed above the color scale bar; Range Code is displayed below the color
scale bar as “S##”. Refer to Table 5–7 for the code range.