Adding Color
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
5–44
CFM Mode Sub-Menu Page 5
Adjust
for Ref
CFM
for Ref
Preset
CFM
Timer
Start
Intervl
Intermnt
Scan
Set Up
ECG
1.0s
Figure 5–20. CFM Sub-Menu Page 5
For details on the Contrast Imaging option, refer to
5–25
.
Intermt Scan
Adjust for Ref
Description
ln dual B-Mode (or CFM Mode), the snap
scan image will display or update
according to the value of the Intervl
selection.
NOTE: When Intermittent Scan is on, the
MI increments in 0.1 steps, starting at the
value 0.0.
Only live images (reference image) can be
optimized with Gain and Acoustic Power
controls. It does not effect the snap scan
image.
Benefits
Snap scan image and live image can be
displayed at the same time. The snap
scan image is only stored in Cine memory.
Only reference image can be optimized to
display the desire image as reference.
Values
On or Off.
On or Off.
CFM for Ref
Timer Start
Description
Only live images (reference image) can
optimize the image as B–Mode. The color
area on reference images is erased. It
does not effect the snap scan image.
A clock counter is used to set the contrast
imaging start point.
Benefits
Only reference image can be optimized to
display desire image as reference.
The timer can be set so the elapse time is
displayed to enable a specific interval time
for contrast imaging.
Values
On or Off.
On or Off.