LV Measurement Methods
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
10–12
Bullet Method (cont’d)
Calc Mnemonic
Calc Name
Input Measurements
Formula
LVLd
Left Ventricular Length,
Diastole
one distance
LVLd=d1[cm or mm]
LVLs
Left Ventricular Length,
Systole
one distance
LVLs=d1[cm or mm]
LVAMd
Left Ventricular Area,
Mitral Valve, Diastole
one area (by ellipse,
trace or circle)
LVAMd=a1[cm ^2]
LVAMs
Left Ventricular Area,
Mitral Valve, Systole
one area (by ellipse,
trace or circle)
LVAMs=a1[cm ^2]
EdV
End Diastole Volume
one distance and one
area (by ellipse, trace or
circle)
EdV[ml]=5xLVLd
xLVAMd/6
EsV
End Systole Volume
one distance and one
area (by ellipse, trace or
circle)
EsV[ml]=5xLVLs
xLVAMs/6
SV
Stroke Volume
two distances and two
areas (by ellipse, trace
or circle)
SV[ml]=EdV–EsV
CO
Cardiac Output
two distances and two
areas (by ellipse, trace
or circle) and one 2 beat
time interval
CO[1/min]=SVxHR/1000
EF
Ejection Fraction
two distances and two
areas (by ellipse, trace
or circle)
EF=SV/EdVx100
HR
Heart Rate
(beats/minute)
one 2 beat time interval
HR[BPM]=120 [sec]/ 2
beat time [sec]
PHT
Pressure Half Time
one time interval
PHT=t1[ms orsec]
MVA
Mitral Valve Area
one pressure half time
interval
MVA[cm^2]= 220/PHT
ET
Ejection Time
one time interval
ET=t1[ms or sec]
Table 10–3. Bullet Method Formulas