OB Summary Reports
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
9–31
Biophysical Profile
A biophysical profile source is available to evaluate the well
being of the fetus. The user can rate the fetus with a score of 0,
1 or 2 on the following items:
Movement
Rate fetal physical activity
Tone
Rate fetal tone
Breathing
Rate fetal breathing movement
Fluid
Rate fetal amniotic fluid
Reactive NST
Rate a Reactive Non-Stress Test
A sum of the above 5 items is then displayed as an overall
biophysical profile score.
If any value is entered other than 0, 1 or 2 for each of the 5
items, an error message “Invalid, Input a proper data (0–2)” is
displayed.
HR, PI, RI, S/D (D/S) Ratio and A/B Ratio
The largest portion of page 2 displays the last 3 measurements
and latest/average value for HR, PI, RI, S/D (D/S) Ratio and
A/B Ratio. The last 4 measurements can be taken at 3 different
sites on the fetus.
The factory default for these site fields is 1, 2 and 3. However,
the user can freely edit these site fields on the Report Page.
For some regions, these site fields may be common between
pages 1 and 2. If this is the case, edits on either page are
reflected on the other. These edits are remembered after
power down. The maximum length of the site field is 6
characters.
The comments section can be edited freely using the
alphanumeric keyboard or annotation library. Simply move the
highlight cursor into the comment area and press the Comment
key. The system is now in the annotation mode. When the
highlight cursor leaves the comment area, the system exits the
annotation mode. These comments are separate from those on
page 1.