M–Mode
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
5–79
M-Mode Sub-Menu Page 2
Color
Tag
MID
Preset
M
Tag
Positn
Color
Sweep
Speed
Set Up
ECG
Figure 5–37. M-Mode Sub-Menu Page 2
Sweep Speed
Color
Description
Changes the speed at which the timeline
updates across the display.
NOTE: Time or distance measurements
are not allowed across sweep speed
changes.
Allows for enabling M-Mode image
colorization.
Benefits
Speed up or slow down the timeline in
order to view more or fewer occurrences
over a period of time.
A fast speed shows less cycles but better
transition definition.
A slow speed shows more cycles but less
definition during transitions.
Displaying the gray scale as shades of
color may allow for improved differentiation
between echo levels.
Values
Slow (16 sec), Mid (8 sec) and Fast (4
sec).
Pre-processing function.
On or Off.
The M-Mode color presented is a selection
in the Set Up/Custom Display menu
page 14.
Post processing function.
Affects on
Other Controls
If more events are seen, the timeline
appears smaller; if less events are seen,
the timeline appears larger or more spread
out.