Troubleshooting
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
16–31
Operation Error Message Description (cont’d)
Category
Problem/
Message
Possible
Cause
Possible
Corrective Action
Storage Function
Mismatched System ID.
Insert correct media.
The disk is not a
registered disk. Images
cannot be deleted.
Take disk to system with
proper ID.
The MO is full, change
MO.
There is not enough
space to store additional
images on the disk.
Use a different disk.
Media is locked.
The disk has been write
protected.
Disable the write
protection or use another
disk.
Media is either
unmounted or
unformatted.
The disk is either not
formatted or inserted in
the drive.
Either format the disk or
insert it into the drive.
Too much data. Store
failed.
The hard drive capacity
of the hard drive to store
additional MOD media
search data is full.
Backup MOD media
search data from the
hard drive to a separate
MOD before attempting
image storage.
This function in not
available.
During the execution of
the background storage
process, this message
appears if Ctrl + W,
Record 1 or Record 2
keys are pressed.
Miscellaneous
No function is defined.
No function is defined for
the User Define key that
was pressed.
Use setup menu to
define.
(Y/N)
Operator Decision
Requested
Type in Y for Yes or N for
No
Table 16–4. Operation Error Message Description (cont’d)