Custom Display
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
14–19
Page 3 of 20 (Imaging Parameter 3 – Probe Dependent 3)
Timeline Sweep Speed EC, UP, P, R
Choose the default value for the timeline (M-Mode and Doppler spectrum) sweep speed.
Slow
16 microseconds per pixel
Mid
8 microseconds per pixel
Fast
4 microseconds per pixel
Very-Fast*
2 microseconds per pixel
* Works with Doppler only, not M-Mode or M/D-Modes.
Physio Sweep Speed on B EC, UP, P, R
Choose the default value for the physio signal sweep speed displayed on the B-Mode
image.
Slow
16 microseconds per pixel
Mid
8 microseconds per pixel
Fast
4 microseconds per pixel
M Gain (Delta from B) EC, UP, P, R
Choose the default value for M-Mode gain difference from the B-Mode Gain.
Value from –98 to 98 in 2 digit increments.
M Dynamic Range EC, UP, P, R
Choose the M-Mode dynamic range default value.
Value from 30 to 90 in 6 digit increments.
M Edge Enhance EC, UP, P, R
Choose the M-Mode edge enhancement default level.
Off, Low, Mid or High.
M Image Softener EC, UP, P, R
Choose the M-Mode image softener function default.
On or Off.
M Softener Level EC, UP, P, R
Choose the M-Mode image softener default level.
Low, Mid or High.