Adding Color
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
5–37
CFM Mode Sub-Menu Page 1 (cont’d)
Diag Mode
Color Flow MTI Filter
Description
Allows for the selection of the best color flow display
format. The 4 methods are:
Survey Mode (Svy)—The Survey Mode
is the default Diag Mode display and is
common for each of the 2 programs.
The Survey Mode displays the B/W
B-Mode image and has an adjustable,
full depth CF wedge that can be steered
with the Trackball.
Survey/Detail Mode (SvyDtl)—The
Survey/Detail Mode displays a high
resolution color flow window that is half
the depth of the B-Mode image. The CF
wedge is adjustable horizontally and
ÇÇ
ÇÇ
ÇÇ
ÇÇ
ÇÇ
ÇÇ
Filters out low flow velocity
color. This minimizes
motion artifacts caused
from breathing.
edge s adjustab e o
o ta y a d
vertically and can be steered within the
full B-Mode image with the Trackball.
Map Mode (Map)—In the Map Mode,
the B-Mode image and the color flow are
the same size. The Map Mode offers
higher CF frame rates and a full size
adjustable CF wedge.
Map/Detail Mode (MapDtl)—The Detail
Mode shows a high resolution, high
frame rate color flow window that is half
of the B-Mode depth. The B-Mode and
the CF are the same adjustable angle.
ÇÇ
ÇÇ
ÇÇ
ÇÇ
ÇÇ
ÇÇ
Benefits
Gives the operator a choice of CFM displays to optimize
scanning.
Filters out the lowest
velocities (colors).
Values
4 sequences are available in the Set Up/Custom Display
menu page 17:
Survey>Map>Map/Detail
Survey>Survey/Detail
Survey>Survey/Detail>Map
Survey/Detail>Map/Detail
Pre-processing function.
Values are Low, M1, M2,
M3, M4 and High.
Pre-processing function.
Affects on
Other Controls
Display frame rate could be improved by minimized gray
scale B-Mode image size or CFM window size.