System Care and Maintenance
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
16–14
Cleaning the system (cont’d)
Monitor (cont’d)
Use a soft, folded cloth and a glass cleaner solution. Apply the
glass cleaner to the cloth. Gently wipe the monitor face and
filter.
Do NOT use a glass cleaner that has a hydrocarbon base (such
as Benzene, Methyl Alcohol or Methyl Ethyl Ketone) on
monitors with the filter (anti-glare shield). Prolonged use of
such cleaners will damage the filter (anti-glare shield). Hard
rubbing will also damage the filter.
NOTE: When cleaning the monitor, make sure not to scratch
the monitor.
After cleaning the monitor face and filter, re-install the filter on
the monitor by reversing the removal procedure in Figure 16–1.
Insert the filter clamps and slide them inwards securely.
NOTE: Make sure that the filter is securely fixed by the filter
clamps. These are important to prevent a hazard from the filter
falling. A “click” sound will be heard when the filter clamp is
secured by the lock mechanism.
Operator Controls
To clean the operator control panel:
1.
Moisten a soft, non-abrasive folded cloth with a mild,
general purpose, non-abrasive soap and water solution.
2.
Wipe down operator control panel.
3.
Use a cotton swab to clean around keys or controls.
Use a toothpick to remove solids from between keys
and controls.
NOTE: When cleaning the operator control panel, make sure
not to spill or spray any liquid on the controls, into the system
cabinet, or in the probe connection receptacle.
CAUTION