Probe Overview
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
15–15
Probe Cleaning Process
To clean the probe:
1.
After each use, disconnect the probe from the
ultrasound console and remove all coupling gel from
the probe by wiping with a soft cloth and rinsing with
flowing water.
2.
Wash the probe with mild soap in lukewarm water.
Scrub the probe as needed using a soft sponge, gauze,
or cloth to remove all visible residue from the probe
surface. Prolonged soaking or scrubbing with a soft
bristle brush (such as a toothbrush) may be necessary
if material has dried onto the probe surface.
3.
Rinse the probe with enough clean potable water to
remove all visible soap residue.
4.
Air dry or dry with a soft cloth.
Special Cleaning Instructions for the E721(MTZ) and ERB7:
When cleaning the E721(MTZ) or ERB7 probe, it is important to
be sure that all surfaces are thoroughly cleaned. These probe
have an adjustable two-part handle that must be disassembled
to gain access to all surfaces. To disassemble the handle,
completely remove the handle adjustment screw located
mid-way between the cable entry and probe tip. The two
handle halves and adjustment screw must be thoroughly
cleaned along with the main probe shaft as described earlier in
step 2. After rinsing and drying is completed, the probe handle
can be loosely reassembled for the disinfection process.