Preset Program
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
14–70
Page 1 of 10 (Application)
Body Pattern Probe Mark Preset EC, UP
Choose to activate the preset position of the probe marker graphic.
On or Off.
Body Pattern Package EC, UP
Choose one of the 8 customized body pattern packages to be displayed with the current
application preset.
Body pattern package 1, 2, 3, 4, 5, 6, 7 or 8.
Refer to
14–59 for details on programming the contents of each body pattern package.
Comment Erase at Display Change EC
Choose to erase all comments when a change is made in scan mode.
On or Off.
Comment Home Position Horizontal EC, R
Choose a point on the horizontal axis for the comment cursor home position.
Enter 1 to 56 in 1 digit increments.
Comment Home Position Vertical EC, R
Choose a point on the vertical axis for the comment cursor home position.
Enter 1 to 26 in 1 digit increments.
Comment Clear Key Function EC
Choose to erase all comments with the Clear key or just the line of comments that the
cursor is on.
All
Erase all comments on the screen.
Line
Erase the comments on the cursor line. The second time Clear is pressed, all
comments will be erased.
Trackball Function at Freeze EC
Choose the function assigned to the Trackball when the image is frozen.
No
No function assigned.
Measure
Trackball assigned to the measurement function.
Comment
Trackball assigned to the comment function.
Body
Trackball assigned to the body pattern function.