Mode Measurements
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
7–18
Time Interval
To measure a horizontal time interval:
Press Measurement four times. A “
” cursor with a vertical
dotted line appears when the cursor is in the Doppler spectrum
portion of the display.
Use the Trackball to move the cursor to the measurement start
point.
Press Set to fix the first cursor.
Use the Trackball to move the end cursor to the measurement
end point.
Press Set to complete the measurement.
The time interval between the two cursors is displayed at the
bottom of the screen.