Powering On the System
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
3–17
Powering On the System
Connecting and Using the System
To connect the system to the electrical supply:
1.
Ensure that the wall outlet is of the appropriate type.
2.
Ensure that the power switch is turned off.
3.
Unwrap the power cable. Make sure to allow sufficient
slack in the cable so that the plug is not pulled out of
the wall if the system is moved slightly.
4.
Push the power plug securely into the wall outlet.
To avoid risk of fire, the system power must be supplied from a
separate, properly rated outlet. See
Site Requirements, Before
the system arrives
on
3–4
for rating information.
Under no circumstances should the AC power plug be altered,
changed, or adapted to a configuration rated less than
specified. Never use an extension cord or adapter plug.
To help assure grounding reliability, connect to a “hospital
grade” or “hospital only” grounded power outlet.
220–240 VAC, 1350 VA
Plug and Outlet Configuration
(Europe)
120 VAC, 1350 VA
Plug and Outlet Configuration
(USA)
Figure 3–8. Example Plug and Outlet Configurations
WARNING