Page 8 of 20 (Imaging Parameter 8 – Probe Dependent 8) (cont’d)
PDI Persistence EC, UP, P, R
Choose the default setting for Persistence, which retains highest color pixel value for a
specified period of time or until a higher value is detected.
OFF, SHORT, MID or HIGH.
PDI MR-Flow EC, UP, P, R
Choose the default setting of the MR-Flow option (combination focus in PDI color area).
Limited to LA39, C551 (CAE), C358, E721, 739L and 546L probes only. Must have
ACE-2 option installed.
On or Off.
PDI MR-Flow Combi Number EC, UP, P
Set the number of focal zones to be used for the combination focus for MR-Flow option.
2 or 3.
Page 9 of 20 (Imaging Parameter 9 – Probe Dependent 9)
PFD Color Threshold EC, UP, P, R
Choose a priority level for B-Mode display during PFD mode. The higher the number,
the greater the emphasis on B-Mode verses PDI Mode.
Choose 0 to 100 in 1 digit increments.
PFD Threshold Power EC, UP, P, R
Set the threshold value (%) at which power information will start to be displayed.
Set from 0 to 13 percent.
Example: Setting = 10%; Power values less than 10% will not be displayed. They will
be set to the background color. Only power values over 10% will be displayed as color.
PFD Threshold Pulsation EC, UP, P, R
Set the threshold value (%) at which pulsation information will start to be displayed.
Set from 0 to 100 percent.
Example: Setting = 20%; Pulsation values less than 20% will not be displayed. They
will be set to the background color. Only pulsation values over 20% will be displayed as
color.
PFD Type EC, UP, P, R
Choose the desired type of the algorithm for detection of PFD.
1, 2 or 3.