Using Cine
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
6–8
Using Cine
Introduction
Cine is useful for focusing on images during a specific part of
the heart cycle or to find an image before the patient moved or
breathed.
Cine images are constantly being stored by the system. Cine
memory capacity is 31 (0–30) frames with 128 scan lines. The
Cine Memory stores the most recent data available for playback
or manual review via Cine. The amount of time represented
depends on the system’s frame rate, scan mode, image size
and other parameters.
Timeline data is continually stored at four times the display
width of timeline data (and updates the corresponding B-Mode
images).
View Cine as a continuous loop via Cine Loop or manually
review Cine images frame by frame via the Cine Scroll Control.
Data in Cine is available until new data is acquired. Cine is
stored in the system’s memory and can be transferred to image
memory, video page printer, multi-image camera or optional
optical disk (MOD).
An Extended Cine Memory Option is available to increase Cine
storage capabilities.