B–Mode
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
5–14
Controls (cont’d)
Reverse
Display Format
(Dual)
Description
Controls the horizontal orientation of the
image on the screen. It toggles the
left/right orientations of the image display.
Each scan mode or combinations of
modes can be displayed in dual format
(side by side images). The Left Image
and Right Image keys allow for the dual
format to be displayed.
Accessing/
Changing
Press the Reverse key to toggle image
reverse on or off.
To initiate the dual format display, press
the Left Image key. The current display
mode will be reduced to the left half of the
image area.
Press the Right Image key to activate the
same image format on the right half of the
image area.
NOTE: When the Left Image key and
Right Image key are pressed
simultaneously, while in single B-Mode
imaging, the dual live image mode is
activated.
Benefits
The Reverse key allows for changing the
orientation of the image display without
physically rotating the probe 180
°
.
Allows for viewing two images
side-by-side.
Values
On or Off.
The GE logo at the top of the sector
wedge corresponds to the orientation mark
on the probe body. Pressing the Reverse
key flips the image left/right and switches
the GE logo to correspond to the probe
mark.
System parameter displays relocated for
each format.
Affects on
Other Controls
Other controls only affect the active image.