Custom Display
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
14–48
Page 20 of 20 (Imaging Parameter 20)
3D Swing Speed EC, UP, R
Choose the swing speed of the displayed MIP image.
1/1, 1/2, 1/4, 1/8 or 1/16 with 1/1 being the fastest swing speed.
3D Scan Line Resolution EC, UP
Choose the default setting for 3D-Mode Scan Line Resolution.
Normal or High.
3D Number of Frames in Build EC, UP
Choose the number of frames to be constructed for the final MIP image. The higher the
number the better the 3D resolution but with longer rendering time.
Number from 5 to 21 in 2 frame steps.
3D B Percent of Source Frames EC, UP
Choose the number of source frames to be used for constructing the B/W MIP image.
Number from 1/1, 1/2, 1/3, 1/4 or 1/5.
3D B Projection Technique EC, UP, R
Choose the method to be used in constructing the MIP image.
Min
Minimum Intensity Projection.
Max
Maximum Intensity Projection.
Grad
Backward images are faded.
Inv
Same as Min but the gray levels of the constructed images are inverted.
3D B Gradient Threshold [%] EC, UP
Choose the gray level threshold percentage to be used in the B/W Gradient Mode.
Percentage from 0 to 100% in 2% increments.
3D B Background Level [%] EC, UP
Choose the percentage of the gray level for the background of the B/W MIP images.
Percentage from 0 to 100% in 2% increments.
3D B Frame Line Level [%] EC, UP
Choose the percentage of the gray level to be used for the frame line of the B/W MIP
image.
Percentage from 0 to 100% in 2% increments.