Recording Images
LOGIQ 500 Basic Users Manual
2276612–100 Rev. 0
13–24
Related Preset Parameters
Two selections must be made in the Set Up/Preset Program
Sub-Menu page 2 that affect image archiving.
Image Archive Compression:
S
No is no DEFF format image compression.
S
Yes is for DEFF format image compression. Storage time
for a compressed image is more than double.
NOTE: If image compression is used, images saved by the
LOGIQ
500 cannot be read on other TIFF readers, DEFF
devices or the LOGIQ
700. However, they can be read on
the LOGIQ
400.
B/W Image Archive & Color Image Archive:
S
B/W is for storing in black and white format (less storage
space required).
S
Color is for storing in color format (more storage space
required).
NOTE: An External Video Image is recognized as a color
image and will use the Color Image Archive preset.
System ID Entry/Display
The LOGIQ
500 system ID number is assigned at the factory.
A System ID should be unique for a system. This number is
used by the system when initializing a MOD (image archive
option) for image storage.
If there is more than one system at a facility, the ID numbers will
be different. A scanner will write to a MOD if the media was
initialized on a different scanner (different ID number).
However, images can only be deleted on the system that
initialized the MOD.
NOTE: If there is more than one LOGIQ
400 or LOGIQ
500
in a facility or group of hospitals, each system should have it’s
own unique System ID number, unless it is necessary to delete
MOD images on any system.
.
.
.